Visit other Fluke companies for calibration equipment and services in electrical, power, pressure, flow, RF, and more.
|
1523/1524 Reference ThermometersThis handheld digital Reference Thermometer from Fluke is a precision, portable, electronic thermometer with trending and logging capability for NIST-traceable calibratin and high accuracy. The 1523 reads a PRT, RTD, thermocouple or thermistor and the 1524 can read two probes with simultaneous measurements. Measure, graph and record three sensor types with one tool
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
1523 single-channel thermometer |
1524 two-channel thermometer |
![]() |
![]() |
| 1 RS-232 serial interface connector. For PC communications, uploading and downloading data from memory and from the probe INFO-CON connectors. | 1 RS-232 serial interface connector |
| 2 Sensor connector (PRT, thermocouple, or thermistor) | 2 Channel 1 sensor connector (PRT, thermocouple, and thermistor) |
| 4 External power adapter connection, for continuous use without changing batteries. Alternatively, 3 AA batteries will last more than 20 hours in the field. | 3 Channel 2 sensor connector (PRT, and thermistor) |
| 4 External power adapter connection | |
| The 1523 Reference Thermometers are versatile single-channel thermometers that measure, graph and record three sensor types with one tool. Support for PRTs/RTDs, thermocouples, and thermistors provide flexibility to choose the right probe for the job. | The new 1524 Reference Thermometers help you do twice the work in half the time. Two channels and three sensor types and high-speed measurement make you more productive and make model 1524 the one reference thermometer you need to own. It has all the features of the 1523, and it’s a data logger too. A real-time clock and memory for 15,000 time and date stamped measurements mean everything you are going to need is in this package. Log up to three times per second, or once every hour or other options in between. Download the data to a PC for analysis when you need it. |
|
1523 |
1524 |
|
|
Input channels |
1 |
2 |
|
Resolution |
PRTs and thermistors: 0.001° |
|
|
Logging |
25 readings with statistics |
25 readings with |
|
Sample interval (normal) |
1 second |
1 second |
|
Typical sample interval (fast mode)* |
0.3 seconds |
|
|
Sensor types |
PRTs, RTDs, thermistors, and thermocouples |
|
|
Thermocouple types |
C, E, J, K, L, M, N, T, U, B, R, S |
|
|
Statistics |
Maximum, minimum, average, standard deviation |
|
|
Trending |
Scale: ± 10 °C (18 °F), ± 1 °C (1.8 °F), ± 0.1°C
(0.18 °F), |
|
|
Power requirements |
3 AA alkaline batteries, 12 V dc universal power supply |
|
|
Size (HxWxD) |
96 mm x 200 mm x 47 mm (3.75 in x 7.9 in x 1.86 in) |
|
|
Weight |
0.65 kg (1.4 lb) |
|
|
Computer interface |
RS-232, 9940 I/O ToolKit software included |
|
|
Safety |
EN61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 |
|
| Environmental conditions for best accuracy: 13 °C to 33 °C (55.4 °F to 91.4 °F) | ||
|
Millivolt range and accuracy |
–10 mV to 75 mV, ± (0.005 % + 5 µV) |
|
|
Internal reference junction compensation |
± 0.2 °C (± 0.36 °F) |
|
|
Resistance range and accuracy |
0 W to 400 W ± (0.004 % + 0.002 W) |
|
|
Temperature coefficient, voltage: |
± (0.001 %/°C + 1 mV/°C) |
|
|
Temperature coefficient, resistance: |
0.0008 %/°C + 0.0004 W (0 W to 400 W) |
|
|
Excitation current, resistance |
1 mA (0 W to 400 W) |
|
|
*See technical manual for sample interval details by sensor type and number of inputs. |
||
Equivalent temperature accuracies for selected sensors derived from primary specifications (Ω, mV)
|
Type |
Range |
Measure accuracies |
|
K |
–200 °C to 0 °C |
± 0.61 °C |
|
0 °C to 1370 °C |
± 0.24 °C |
|
|
R |
–20 °C to 0 °C |
± 1.09 °C |
|
0 °C to 500 °C |
± 0.97 °C |
|
|
500 °C to 1750 °C |
± 0.49 °C |
|
|
S |
–20 °C to 0 °C |
± 1.05 °C |
|
0 °C to 500 °C |
± 0.95 °C |
|
|
500 °C to 1750 °C |
± 0.56 °C |
|
|
Accuracies are based on internal reference junction compensation. Refer to technical manual for greatly improved accuracies using external reference junctions. |
||
|
Accuracy ± °C 4 (K=2) |
|
± 0.011 at –100 °C |
|
± 0.015 at 0 °C |
|
± 0.019 at 100 °C |
|
± 0.023 at 200 °C |
|
± 0.031 at 400 °C |
|
± 0.039 at 600 °C |
|
Resolution: 0.001°C (0.001°F) |
Accuracy ± °C (K=2) |
|
± 0.002 at 0 °C |
|
± 0.003 at 25 °C |
|
± 0.006 at 50 °C |
|
± 0.014 at 75 °C |
|
± 0.030 at 100 °C |
|
Resolution: 0.001 °C (0.001 °F) |
|
Based on a 10 kW (at 25 °C) thermister with a beta value of 4000 W. See technical manual for details. |
|
Temperature |
5616-12 |
5615-6 |
5627A-12 |
5610-9 |
|
-200 °C (-328 °F) |
0.014 |
0.025 |
0.027 |
n/a |
|
0 °C (32 °F) |
0.021 |
0.021 |
0.049 |
0.009 |
|
100 °C (212 °F) |
0.027 |
0.028 |
0.065 |
0.030 |
|
300 °C (572 °F) |
0.040 |
0.043 |
0.103 |
n/a |
|
420 °C (788 °F) |
0.050 |
n/a |
0.130 |
n/a |
|
Includes readout accuracy, |
||||
| 1523* | Reference Thermometer, Handheld, 1 Channel |
| 1524* | Reference Thermometer, Handheld, 2 Channel, Data Logger |
| 1523-P1 | 1523 Bundled with 5616 PRT [–200 °C to 420 °C (-328
°F to 788 °F), NIST Traceable Probe Calibration, 100 ohm, 6.35 mm x 305 mm
(1/4 in x 12 in)], Universal TC INFO-CON Connector, TPAK, and Case |
| 1523-P2 | 1523 Bundled with 5628 PRT [–200 °C to 660 °C (-328
°F to 1220 °F), Accredited Probe Calibration, 25 ohm, 6.35 mm x 305 mm
(1/4 in x 12 in)], Universal TC INFO-CON Connector, TPAK, and Case |
| 1523-P3 | 1523 Bundled with 5627A PRT [–200 °C to 420 °C (-328
°F to 788 °F), Accredited Probe Calibration, 100 ohm, 6.35 mm x 305 mm
(1/4 in x 12 in)], Universal TC INFO-CON Connector, TPAK, and Case |
| 1524-P1 | 1524 Bundled, with 5616 PRT, Universal TC INFO-CON Connector, TPAK, and Case |
| 1524-P2 | 1524 Bundled with 5628 PRT, Universal TC INFO-CON Connector, TPAK, and Case |
| 1524-P3 | 1524 Bundled with 5627A PRT, Universal TC INFO-CON Connector, TPAK, and Case |
| *Requires an optional probe | |
Calibration options |
|
| 1523-CAL | 1523 Accredited Calibration |
| 1524-CAL | 1524 Accredited Calibration |
| 1929-2 | System Verification, PRT with Readout, Accredited |
| 1929-5 | System Verification, Thermistor with Readout, Accredited |
| 1930 | System Calibration, PRT with Readout, Accredited |
| 1935 | System Calibration, Thermistor with Readout, Accredited |
| 1925-A | Accredited Thermistor Calibration, 0 °C to 100 °C (23 °F to 212 °F) |
Included accessories |
|
| NIST traceable certificate of calibration, users guide, CD-ROM (contains technical manual), 12 V dc universal power supply, RS-232 cable, 9940 I/O ToolKit software | |
Optional accessories |
|
| 5610-9-P | Probe, Precision Thermistor, Stainless Steel, 3.18 mm x 228.6 mm
(1/8 in x 9 in), 0 °C to 100 °C (32 °F to 212 °F), NIST traceable calibration |
| 5615-6-P | Probe, PRT, 100 ohm, 4.76 mm x 152.4 mm (3/16 in x 6 in), –200
°C to 300 °C (–328 °F to 572 °F), Accredited Calibration |
| 5609-9BND-P | Probe, PRT, 100 ohm, 6.35 mm x 381 mm (1/4 in x 15 in), 90° bend
at 9 inches, –200 °C to 660 °C (–328 °F to 1220 °F), Requires Calibration (i.e. 1924-4-7) |
| FLK80P1 | 80PK-1, Probe and adapter (2373-LTC), Thermocouple, Beaded Type K, –40 to 260 °C |
| FLK80P3 | 80PK-3A, Probe and adapter (2373-LTC), Thermocouple, Surface Measurement Type K, 0 to 260 °C |
| 9935-S | Software, LogWare II, Single User |
| 1523-CASE | Case, 1523/1524 Readout and Probe Carrying ![]() |
| FLUKETPAK | TPAK, Meter Hanging Kit ![]() |
| 2373-LPRT | Adapter, Lemo to Mini Grabbers
(4-wire) ![]() |
| 2373-LTC | Adapter, Lemo to Universal TC (TC) ![]() |
| 2373-L21 | Adapter, Lemo to INFO-CON (allows the 1523/1524 to read probes terminated with the 1521/1522 INFO-CON) |
| 2384-P | INFO-CON Connector, PRT (Gray Cap), Spare |
| 2384-T | INFO-CON Connector, TC (Blue Cap), Spare |
| 2525-156 | Isolator, RS-232 Optical 12 V dc (1523/1524), 115 V |
| 2523-256 | Isolator, RS-232 Optical 12 V dc (1523/1524), 230 V |
The following items are available for download in Adobe Acrobat (.pdf) format and require Adobe Reader or Acrobat. Click on the link to view in your browser or right click and “Save Target As...” to download to your computer.
Note: Users Guides are available for download in separate language versions. Visit our manual page or go directly to our ftp site area - ftp.hartscientific.com/manuals
Use our feedback form for more information.
Click to request a quote, catalog, or a consultation. Or call us at 1-800-438-4278.